Probes and Biopsy 13-15
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
ELC13-4s
-20℃~55
20%~95%RH (no condensation) 700 hPa ~ 1060 hPa
DE11-3Ws
-10℃~50
20%~95%RH (no condensation) 700 hPa ~ 1060 hPa
L13-3Ns
-20℃~55
20%~95%RH (no condensation) 700 hPa ~ 1060 hPa
L14-3Ws
-20℃~55
20%~95%RH (no condensation) 700 hPa ~ 1060 hPa
L16-4Hs
0℃~60
30%~95%RH (no condensation) 700 hPa ~ 1060 hPa
1. To prevent the probe from being damaged, DO NOT store it in locations where it may be exposed
to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. When the transducer is sent to MINDRAY Customer Service Department or sales representative for
repair, be sure to disinfect it and keep it in the carrying case to prevent infection.
3. Sterilize the carrying case as necessary.
13.1.7 Sensor (Fusion Imaging) Support Cleaning and
Disinfection
Before performing the fusion imaging exam, clean and disinfect (or sterilize) the fusion sensor support
as required.
CAUTION:
1. When performing cleaning and
disinfection of the sensor support to
prevent infection, wear sterile gloves.
2.
After disinfection, rinse the sensor support thoroughly with
sterile water to remove all chemical residues. Chemical residues
may be harmful to the human body.
3
No cleaning and disinfecting may result in the sensor support
becoming a source of infection.
The efficacy of disinfectants and sterilizing solutions is not
guaranteed by MINDRAY. Contact the manufacturers for
information on the activity of the products.
Cleaning
1. Disconnect the sensor support from the system.
2. Wear sterile gloves to prevent infection.
3. Wash the sensor support (fusion) with clean water or soapy water to remove all the foreign matters,
or, wipe the sensor support with a soft ethyl carbamate sponge. Avoid using a brush, because it may
damage the transducer.
4. Dry the sensor support using a sterile cloth or gauze after rinsing. Do not dry the sensor support by
heating it.