13-16 Probes and Biopsy
Disinfecting with Sprays or Wipes
CAUTION:
Use protective eyewear when disinfecting using sprays.
1. Wear sterile gloves to prevent infection.
2. After you have finished cleaning, spray or wipe the sensor support (fusion) with a disinfectant. Follow
the disinfectant manufacturers recommended contact time and mode.
3. Remove any residue with a water-moistened soft cloth on the sensor support.
4. Wipe off water on the sensor support using sterile cloth or gauze after washing.
Disinfecting by Immersion
1. Wear sterile gloves to prevent infection.
2. Clean the sensor support before disinfecting it. MINDRAY recommends the following solutions to
disinfect the sensor support (fusion).
Refer to the instructions provided by the chemical manufacturer concerning concentration of
the disinfectant solution, method of disinfection and dilution and cautions during use.
Soak the sensor support into the disinfectant solution for the shortest time the manufacturer
recommends (for example, the shortest time recommended by the manufacturer for soaking
Cidex OPA is 12 minutes).
Follow local regulations when selecting and using the disinfectant.
3. Rinse the sensor support with plenty of sterile water (about 2 gallons) for at least 1 minute to remove
all chemical residues on it. Or, follow the rinsing method recommended by the disinfectant
manufacturer to rinse the sensor support.
4. Wipe off the water on the sensor support with sterile cloth or gauze after rinsing it.
Compatible Disinfectants
Manufacturer Trade Name Procedures Type
Metrex MetriZyme
Please refer to the instructions
provided by the manufacturer of
the solution for details.
Solution
ASP
Cidex Activated
Glutaraldehyde
Solution
Please refer to the instructions
provided by the manufacturer of
the solution for details.
Solution
ASP Cidex OPA
Please refer to the instructions
provided by the manufacturer of
the solution for details.
Solution
Ecolab Inc. Ster-Bac
Please refer to the instructions
provided by the manufacturer of
the solution for details.
Solution
Advanced Ultrasound
Solutions Inc.
SONO Ultrasound
Wipes
Please refer to the instructions
provided by the manufacturer of
the solution for details.
Wipe
Professional
Disposables
International, Inc.
Sani-Cloth
Please refer to the instructions
provided by the manufacturer of
the solution for details.
Wipe