Probes and Biopsy 13-17
13.1.8 Storage and Transportation for Sensor Support
(Fusion)
When all examinations for the day have been completed, confirm that the sensor support is in good
condition. After disinfecting the sensor support, confirm that the sensor support is in good condition and
stored in a suitable place.
1. To prevent the sensor support from being damaged, DO NOT store it in locations where it may be
exposed to:
Direct sunlight
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. Store and transport sensor support under following ambient conditions:
Ambient temperature: -20°C ~ 55°C
Relative humidity: 20% ~95% (no condensation)
Atmospheric pressure: 700 hPa ~1060 hPa
13.1.9 Sensor (Fusion Imaging) Cleaning and Disinfection
CAUTION:
1. When performing cleaning and
disinfection of the sensor to prevent
infection, wear sterile gloves.
2.
After disinfection, rinse the sensor thoroughly with sterile water
to remove all chemical residues. Chemical residues may be
harmful to the human body.
3
No cleaning and disinfecting may result in the sensor becoming
a source of infection.
4
The efficacy of disinfectants and sterilizing solutions is not
guaranteed by MINDRAY. Contact the manufacturers for
information on the activity of the products.
Disinfecting with Sprays or Wipes
CAUTION:
Use protective eyewear when disinfecting using sprays.
1. Wear sterile gloves to prevent infection.
2. Spray or wipe the sensor (fusion) with a disinfectant. Follow the disinfectant manufacturers
recommended contact time and mode.
3. Remove any residue with a water-moistened soft cloth on the sensor.
4. Wipe off water on the sensor using sterile cloth or gauze after washing.