Trio™ Operating Instructions 0070-10-0666-01 2 - 73
Operations Alarms
RESP ALARM SETUP Menu Selections
FIGURE 2-53 RESP Alarm Setup Menu
ALM SEL
(Alarm Selection)
Choose RESP ALARM SETUP.
ALM Allows the user to turn the alarm ON or OFF. Choose ON to
enable the alarm; choose OFF to disable the alarm. If the alarm is
set to OFF, the alarm OFF symbol will display within the
parameter tile. The alarm is activated when the measurement
exceeds the high or low limits.
ALM PRIORITY Allows the user to select the priority of the alarm. Choices are 1,
2, and 3. Priority 1 alarms are considered the most serious.
ALM PRINT Allows the user to enable or disable automatic printing during a
physiological alarm condition. Choose ON to enable printing
upon alarm. Choose OFF to disable printing upon alarm. See
“Recorder (Optional)” on page 2-82 for details.
ALM HI Allows the user to set the upper limit of the RESP alarm. (See
“Respiration Rate Alarm Limits” on page 2-30.)
ALM LO Allows the user to set the lower limit of the RESP alarm. (See
“Respiration Rate Alarm Limits” on page 2-30.)