Specifications Appendix
6 - 2 0070-10-0666-01 Trio™ Operating Instructions
6.1.2 Safety Designations
 
EN 12470-4:2000 Clinical thermometers - Part 4: Performance 
of Electrical Thermometers for Continuous 
Measurement
EN 1441:1997 Medical Devices - Risk Analysis
EN ISO14971:2000 Medical Devices-Application of Risk 
Management Analysis to Medical Devices
Type of protection against electric shock -Class 1 with internal electric power source. 
Where the integrity of the external protective 
earth (ground) in the installation or its 
conductors is in doubt, the equipment will be 
operated from its internal electric power 
source (batteries).
Degree of protection against electric shock: - ECG and IBP - Type CF defibrillation 
protected
- NIBP - Type BF defibrillation protected
- SpO
2
 and Temperature - Type BF
- Monitor - Type B
Supply Connection: 100 – 240 Volt
50 or 60 Hz
0.8 – 0.4 Amps
12 VDC Sealed Lead Acid Internal Battery or 
11.1 VDC Lithium Ion Internal Battery
Mode of Operation: Continuous
Protection Against Hazards of Explosion: Not protected (Ordinary)
Protection Against Ingress of Liquid's Not protected (Ordinary) - IPX0 per IEC 529
Degree of electrical connection between 
equipment and patient -
Equipment designed for direct electrical and 
non-electrical connection to the patient
Degree of Mobility: Transportable