Default Configurations Defaults
3 - 6 0070-10-0666-01 Trio™ Operating Instructions
3.1.2 User Default Configuration
All items listed in the preceding table can be customized by the user and saved as a unique 
User Default Configuration as follows:
1. Select SAVE CURRENT from the MONITOR SETUP menu. The confirmation dialog 
box in FIGURE 3-3 is displayed.
2. Select YES to save the new user-defined configuration, replacing the existing user-
defined configuration for the current patient size. Select NO to cancel the task.
FIGURE 3-3 Confirmation Dialog Box 
NOTE: At initial startup of the Trio, the User Default Configuration is 
identical to the Factory Default Configuration.
NOTE: Selecting a new patient size automatically activates the User 
Default Configuration for that patient size.
3.1.3 Current Configuration
The Current Configuration consists of the current settings of the items listed in the 
preceding table. When the monitor is powered OFF, these settings (and any trend data) are 
maintained for 2 hours. When the monitor remains OFF for more than 2 hours (or when a 
patient is discharged), any trend data is deleted. When the monitor is powered back ON (or 
after a patient is discharged), the Current Configuration is replaced by one of two 
configurations in the following hierarchical order:
1. The User Default Configuration (if it has been saved).
2. The Factory Default Configuration (if no User Default Configuration has been 
saved).