Trio™ Operating Instructions 0070-10-0666-01 3 - 5
Defaults Default Configurations
IBP Scale RAP High 40
IBP Scale LAP High 40
IBP Scale ICP High 40
IBP Scale ART Low 0
IBP Scale PA Low 0
IBP Scale CVP Low 0
IBP Scale RAP Low 0
IBP Scale LAP Low 0
IBP Scale ICP Low 0
IBP Scale ART Value 75
IBP Scale PA Value 50
IBP Scale CVP Value 20
IBP Scale RAP Value 20
IBP Scale LAP Value 20
IBP Scale ICP Value 20
IBP Ch1 Name ART
IBP ART_Sys Alarm Limit Hi 180 150
IBP ART_Sys Alarm Limit Lo 80 70
IBP ART_Mean Alarm Limit Hi 100 80
IBP ART_Mean Alarm Limit Lo 40 30
IBP ART_Dia Alarm Limit Hi 100 80
IBP ART_Dia Alarm Limit Lo 50 40
IBP PA_Sys Alarm Limit Hi 180 150
IBP PA_Sys Alarm Limit Lo 80 70
IBP PA_Mean Alarm Limit Hi 100 80
IBP PA_Mean Alarm Limit Lo 40 30
IBP PA_Dia Alarm Limit Hi 100 80
IBP PA_Dia Alarm Limit Lo 50 40
IBP CVP_Mean Alarm Limit Hi 10
IBP CVP_Mean Alarm Limit Lo 0
IBP RAP_Mean Alarm Limit Hi 10
IBP RAP_Mean Alarm Limit Lo 0
IBP LAP_Mean Alarm Limit Hi 10
IBP LAP_Mean Alarm Limit Lo 0
IBP ICP_Mean Alarm Limit Hi 10
IBP ICP_Mean Alarm Limit Lo 0
CONFIGURATION ITEM
FACTORY
DEFAULTS
ADULT
(Patient size)
FACTORY
DEFAULTS
PEDIATRIC
(Patient size)
FACTORY
DEFAULTS
COMMON