Page #119 background imageLoading...
Page #119 background image
Default Configurations Defaults
3 - 4 0070-10-0666-01 Trio™ Operating Instructions
SpO2 Sweep Speed 25 mm/s
SpO2 Beep Volume Medium
SPO2 SENSOR OFF AUDIO ON
SpO2 Average Time (Masimo
only)
8 second
SpO2 Sensitivity Mode Normal
Sat Seconds (Nellcor only) OFF
SpO2 Alarm Limit Hi OFF (for all languages)
SpO2 Alarm Limit Lo 85
Pulse Alarm Limit Hi 150 175
Pulse Alarm Limit Lo 45 70
NIBP SETUP
NIBP Alarm OFF
NIBP Alarm Print OFF
NIBP Alarm Priority 2
NIBP Units mmHg (for all languages)
NIBP Interval OFF
NIBP Display Time-out 15 minutes
NIBP Sys Alarm Limit Hi 180 150
NIBP Sys Alarm Limit Lo 80 70
NIBP Mean Alarm Limit Hi 100 80
NIBP Mean Alarm Limit Lo 40 30
NIBP Dia Alarm Limit Hi 100 80
NIBP Dia Alarm Limit Lo 50 40
TEMP SETUP
Temp Alarm OFF
Temp Alarm Print OFF
Temp Alarm Level 2
Temp Unit °C (for all languages)
Temp T1 Alarm Limit Hi 40 °C or 104 °F
Temp T1 Alarm Limit Lo 30 °C or 86 °F
IBP SETUP
IBP Alarm OFF
IBP Alarm Print OFF
IBP Alarm Priority 2
IBP Units mmHg (for all languages)
IBP Sweep Speed 25 mm/s
IBP SCALE ADJUST
IBP Scale ART High 150
IBP Scale PA High 100
IBP Scale CVP High 40
CONFIGURATION ITEM
FACTORY
DEFAULTS
ADULT
(Patient size)
FACTORY
DEFAULTS
PEDIATRIC
(Patient size)
FACTORY
DEFAULTS
COMMON

Table of Contents

Other manuals for Mindray Trio

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Trio and is the answer not in the manual?

Mindray Trio Specifications

General IconGeneral
BrandMindray
ModelTrio
CategoryMedical Equipment
LanguageEnglish

Related product manuals