Trio™ Operating Instructions 0070-10-0666-01 6 - 19
Appendix Physical Specifications
6.3.4.4 VGA Output
(model numbers 0998-00-0600-0XXXX and 0998-00-0600-2XXXX)
Trio Monitors bearing model numbers 0998-00-0600-0XXXX or 0998-00-0600-2XXXX are 
equipped with a a 15-position D-Shell VGA output connector. The VGA output connector 
provides connectivity to a medical grade remote repeater display. Connection to this port 
shall be made with the monitor power OFF. The monitor shall be powered ON after 
powering ON the remote display.
Working mode: 640X480 resolution, 16-color, APA mode
Signal: Analog RGB - 0.7 Vpp/750 ohm, Horizontal/Vertical - TTL Positive/Negative
Connector type: 15-position D-subminiature receptacle
6.3.4.5 Ethernet
The Trio monitor has a standard Ethernet port for downloading software upgrades.
6.3.5 Power Supply
6.3.5.1 Power Source
The Trio monitor will auto-select its power source from those available. The monitor uses the 
following priority in choosing the power source:
1. AC Mains Power
2. Internal battery power
The monitor will operate from AC Mains power with or without the internal battery installed.
6.3.6 AC Mains Power Source
6.3.7 Battery Power
Connector type: RJ-45
Input Voltage: 100 – 240 VAC (+/-10%)
Line Frequency: 50 or 60 Hz (+/-3 Hz)
Current: 0.8 – 0.4 Amps
The maximum number of installed batteries: 1
Time to Shutdown from Low Battery: >10 minutes but < 20 minutes after 