Alarms Operations
2 - 64 0070-10-0666-01 Trio™ Operating Instructions
General Alarms
General alarms are alerts that cannot be classified as physiological or technical, but that 
require attention. General alarms are not related to patient medical condition.
2.5.2 Alarm Priorities
Alarms in each category (Physiological, Technical or General) are classified based on their 
degree of severity. The priority determines the frequency of audio and visual alerts. There are 
two (2) visual alerts: the alarm light and the alarm-related message on the display. For an 
ECG Lead Off alarm, the alarm will sound according to the ECG priority setting.
1. Priority One (1) Alarm: This alarm requires an urgent and immediate response from 
the clinician. The numeric value of the alarming parameter will flash at a constant rate in 
black text on a red background. The alarm light will flash red with high frequency. The 
audio alert cycles once every ten (10) seconds. 
2. Priority Two (2) Alarm: This alarm requires a serious response and consideration by 
the clinician. The numeric value of the alarming parameter will flash at a constant rate in 
black text on a yellow background. The alarm light flashes amber with a slow frequency. 
The audio alert cycles once every 25 seconds.
3. Priority Three (3) Alarm: This alert should be attended to as soon as possible by the 
clinician. The numeric value of the alarming parameter displays in black text on a yellow 
background and does not flash. The alarm light will be amber and does not flash. The 
audio alert cycles once every 25 seconds. 
All physiological alarms can be changed by the clinician. Technical alarm priorities are 
preset in the system and can not be changed. The alert message for technical alarms displays 
on a yellow background. To change the parameter alarm priorities, open the individual 
PARAMETER SETUP menus and select the priority desired. Parameter alarm priorities may 
also be adjusted in the ALARM SETUP menu.
NOTE: When alarms of different priorities occur simultaneously, the 
monitor will display the alarm with the highest priority.
2.5.3 Alarm Setup Menus
The Alarm Setup Menus enable the customization of the COMMON ALARM SETUP as 
well as the ALARM SETUPs for the following individual parameters: HR, SPO2, NIBP, 
IBP, RESP, and TEMP.
2.5.3.1 Accessing the Alarm Setup Menus
1. Using the Navigator™ Knob, select ALARM SETUP from the MONITOR SETUP 
menu. The default ALARM SETUP menu is the COMMON ALARM SETUP as shown 
in FIGURE 2-47.
2. Use the ALM SEL Drop-Down menu to access the ALARM SETUP menu for a specific 
parameter. The choices are: HR ALARM SETUP, SPO2 ALARM SETUP, NIBP 
ALARM SETUP, IBP ALARM SETUP, RESP ALARM SETUP, and TEMP ALARM 
SETUP. The selected ALARM SETUP menu will determine the other available menu 
selections.
NOTE: For more information on parameter alarms and instructions 
for setting these alarms, refer to the parameter-specific 
section of this Operator's Manual.