Physical Specifications Appendix
6 - 22 0070-10-0666-01 Trio™ Operating Instructions
6.3.11 Environmental and Safety Characteristics
Storage Temperature -20 ºC to +60 ºC
Operating Temperature +5 ºC to +40 ºC
Storage Humidity 10% to 95%, non-condensing 
Operating Humidity 15% to 95%, non-condensing 
Storage Altitude (-1000 to 20,000 feet ASL)
1050 hPa to 466 hPa
(788 mmHg to 349 mmHg)
Operating Altitude (-1000 to 9,889 feet ASL) 
1050 hPa to 700 hPa
(788 mmHg to 525 mmHg)
Shipping: The monitor meets the requirements of ISTA 
shipping procedure 1A for containerized 
product, when packed in designated 
packaging. The shipping carton is marked 
with the permissible storage conditions for 
temperature, altitude, and humidity.
Shock: Remains operational within specification 
after exposure to100 g, 6 ms, half sine, 
shock pulse tested per IEC 68-2-27.
Vibration: Remains operational within specification 
after exposure to the following Sinusoidal 
and Random Vibration
Sinusoidal Vibration Per IEC 68-2-6
1 g or 0.07 mm, 57 – 62 Hz crossover 
frequency 10 to 500 Hz, 10 sweep cycles in 
each axis
Random Vibration Per IEC 68-2-34
0.02 g2/Hz
20 – 500 Hz
Low degree of reproducibility 9 minutes per 
axis