Trio™ Operating Instructions 0070-10-0666-01 6 - 23
Appendix Physical Specifications
Electromagnetic Compatibility
The Trio meets the requirements of IEC 60601-1-2:2001/EN 60601-1-2:2001.
NOTE: The Trio needs special precautions regarding EMC and needs 
to be installed and put into service according to the EMC 
information provided below.
NOTE: Portable and mobile RF communications equipment can 
affect the Trio. See tables 6-1 through 6-4 that follow.
Drop: Meets the requirements specified by ECRI PB-
296 892 section AIII 3.3 for Class 3 
devices. 
Impact Meets the requirements specified by ECRI PB-
296 892, section AIII 3.2 for Class 3 
devices. 
TABLE 6-1
GUIDANCE AND DECLARATION - ELECTROMAGNETIC EMISSIONS
The Trio is intended for use in the electromagnetic environment specified below. The customer or the 
user of the Trio should assure that it is used in such an environment.
EMISSIONS 
TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions 
CISPR 11
Group 1 The Trio uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely 
to cause any interference in nearby electronic equipment.
RF emissions 
CISPR 11
Class A The Trio is suitable for use in all establishments other 
than domestic establishments and those directly 
connected to the public low-voltage power supply 
network that supplies buildings used for domestic 
purposes
Harmonic 
emissions IEC 
61000-3-2
Class A
Voltage 
fluctuations/
Flicker 
emissions IEC 
61000-3-3
Complies