Patient Parameter Specifications Appendix
6 - 10 0070-10-0666-01 Trio™ Operating Instructions
6.2.6.1 Systolic Pressure Measurement
6.2.6.2 Diastolic Pressure Measurement
*Adult and Pediatric
Blood pressure measurements determined with this device are equivalent to those obtained
by a trained observer using the cuff/stethoscope auscultation method, within the limits
prescribed by ANSI/AAMI SP-10:1992, Electronic or automated sphygmomanometers.
NOTE: Mean Arterial Pressure (MAP) is defined as:
Mean Pressure 1 = Mean Pressure determined from the
oscillometric profile
Mean Pressure 2 = (2*diastolic + systolic) / 3
Mean Pressure Displayed = (Mean Pressure 1 + Mean
Pressure 2) / 2
6.2.6.3 Static Pressure Measurement
6.2.6.4 Heart Rate from NIBP
Accuracy*: Mean error is less than ± 5 mmHg, Standard
Deviation is less than ± 8 mmHg
Range:
ADULT MODE PEDIATRIC MODE
40 to 255 mmHg 40 to 200 mmHg
Accuracy*: Mean error is less than ± 5 mmHg, Standard
Deviation is less than ± 8 mmHg
Range:
ADULT MODE PEDIATRIC MODE
10 to 210 mmHg 10 to 150 mmHg
Range: 0 – 325 mmHg
Static Accuracy: ±3 mmHg over the entire range.
Accuracy: ± 2 bpm or 2%, whichever is greater
Resolution: 1 bpm
Range:
ADULT MODE PEDIATRIC MODE
40 to 240 bpm 40 to 240 bpm