Patient Parameter Specifications Appendix
6 - 14 0070-10-0666-01 Trio™ Operating Instructions
SpO
2
Accuracy
No motion conditions
1,4
:
During motion conditions
2,5
:
SpO
2
Response Time
20 seconds maximum to 95% of final step change of % SpO
2
value from 60 to 95% at
75 bpm. Post averaging time is set at 8 seconds
Low Perfusion Performance
3
Masimo SET SpO
2
Pulse Rate Performance Requirements
Pulse Rate Resolution: 1 bpm
Pulse Rate Range and Accuracy
Update Rate
Update rate is 1 Hz.
PATIENT SIZE SATURATION RANGE
70% to 100% 0 – 69%
Adult Mode ± 2 digits unspecified
Pediatric Mode ± 2 digits unspecified
PATIENT SIZE SATURATION RANGE
70% to 100% 0 – 69%
Adult Mode ± 3 digits unspecified
Pediatric Mode ± 3 digits unspecified
LOW PERFUSION CONDITIONS
PULSE
AMPLITUDE % TRANSMISSION
SATURATION
ACCURACY
PULSE RATE
ACCURACY
> 0.02% > 5% ± 2 DIGITS ± 3 DIGITS
LNOP TC-I ± 4 digits ± 3 digits
LNOP Y-I N/A N/A
PATIENT SIZE PULSE RATE RANGE ACCURACY
NO MOTION
CONDITIONS
1
DURING MOTION
CONDITIONS
2
Adult/Pediatric 25 to 240 bpm ± 3 digits ± 5 digits