Parameter Menus Operations
2 - 28 0070-10-0666-01 Trio™ Operating Instructions
Excessive alarms: 
Heart rate, lead 
fault
Electrodes dry  Repeat skin preparation and apply 
fresh, moist electrodes.
Alarm limits set too close to patient's 
normal heart rate
Readjust alarm limits.
R-wave wrong size Readjust the waveform size or 
move ECG electrodes to optimize 
R-wave.
Excessive patient movement or muscle 
tremor
Reposition electrodes and secure 
with tape if necessary.
Low Amplitude 
ECG Signal
Gain set too low Readjust the ECG wave gain as 
required. Refer to "ECG Lead, Size 
and Filter Settings" on page 2-23 
for instructions on adjusting the 
wave gain setting.
Electrodes dry/old Apply fresh, moist electrodes.
Skin improperly prepped Abrade the skin and repeat skin 
preparation.
This could be the patient's normal 
QRS complex.
Verify with a 12-lead electro-
cardiogram.
Electrode positioned over a bone or 
muscle mass
Reposition electrodes.
No ECG waveform Size not set properly Readjust the ECG wave gain as 
required. Refer to "ECG Lead, Size 
and Filter Settings" on page 2-23 
for instructions on adjusting the 
wave gain setting.
Lead wires and/or patient cable not 
fully inserted into proper receptacle
Check cables for proper connection
Cables or lead wires damaged Check with a continuity tester.
Base Line Wander Patient moving excessively. Secure lead wires and cable to 
patient.
Patient respiration variance. Reposition electrodes
Electrodes dry or loose Repeat skin preparation and apply 
fresh, moist electrodes.
MESSAGE/
PROBLEM REASON SOLUTION