EasyManuals Logo

Mindray Trio Operating Instructions

Mindray Trio
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
Parameter Menus Operations
2 - 38 0070-10-0666-01 Trio™ Operating Instructions
Masimo SET SpO
2
Alarm Limits
NOTE: If the SpO
2
alarm high is set to OFF, the alarm OFF
symbol will display in the SpO
2
parameter tile.
Alarms occurring during the process of SpO
2
measurement include two (2) types:
physiological alarms and technical alarms. Physiological alarms occur when the patient’s
pulse rate or oxygen saturation level is equal to or exceeds set alarm limits. Technical alarms
are any SpO
2
-related alarms, which are not physiological, such as functional failures.
2.4.3.1.2 Masimo SET
®
SpO
2
Troubleshooting
PATIENT SIZE
SpO
2
HIGH
ALARM (%)
SpO
2
LOW
ALARM (%)
PR HIGH
ALARM (bpm)
PR LOW
ALARM (bpm)
Adult (ADU) 80 – 100 [OFF] 50 – 99 [85] 60 – 240 [150] 25 – 120 [45]
Pediatric (PED) 80 – 100 [OFF] 50 – 99 [85] 100 – 240 [175] 25 – 150 [70]
* Factory default values shown in brackets.
*
MESSAGE REASON ACTION
SpO
2
: Sensor Off SpO
2
sensor may be
disconnected from the patient
Place the sensor on the patient.
SpO
2
: No Sensor SpO
2
sensor may be
disconnected from the monitor
or the extension cable
Plug the sensor into the monitor or the
extension cable.
SpO
2
: Interference Noise detected on the pulse
signal prevents pulse
discrimination
Decrease patient motion. Check
sensor.
SpO
2
: Pulse Search Hardware settings are being
adjusted in order to
discriminate a pulse waveform
Wait several seconds for saturation
value to be displayed. If it does not
display, do one of the following:
Change to site where pulse is
stronger if patient is
vasoconstricted.
Change or readjust sensor if loose.
SpO
2
: Low Perfusion Patient perfusion is low Check patient connection and patient
status
SpO
2
: Too Much Light There is too much ambient room
light for the sensor to function
properly
Minimize the room light around the
patient. Check sensor.
SpO
2
: Unrecognized
Sensor
The sensor is not recognized by
the monitor
Replace the sensor with a
recommended authorized sensor
SpO
2
:
Communication Error
The monitor and the SpO
2
modules are not communicating
properly
Power the unit OFF/ON. If problem
persists, notify hospital technician or
Customer Support.
SpO
2
: Board Fault Masimo SET board failed to
operate properly
Notify hospital technician or Customer
Support
SpO
2
: Sensor Fault Defective Sensor Replace sensor
To Purchase, Visit Avobus.com or call 1-800-674-3655

Table of Contents

Other manuals for Mindray Trio

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Trio and is the answer not in the manual?

Mindray Trio Specifications

General IconGeneral
BrandMindray
ModelTrio
CategoryMedical Equipment
LanguageEnglish

Related product manuals