Alarms Operations
2 - 66 0070-10-0666-01 Trio™ Operating Instructions
HR ALARM SETUP Menu Selections
FIGURE 2-48 HR Alarm Setup Menu
ALM SEL
(Alarm Selection)
Choose HR ALARM SETUP.
ALM
NOTE: In the French configuration, the HR 
(Heart Rate) alarm is always ON. It 
cannot be disabled by turning it 
OFF.
Allows the user to turn the HR (Heart Rate) alarm ON or OFF. 
Choose ON to enable the alarm; choose OFF to disable the 
alarm. If the alarm is set to OFF, the alarm OFF symbol   will 
display to the right of ECG on the screen. The HR alarm is 
activated when the heart rate is equal to or exceeds set high or 
low HR values.
ALM PRIORITY Allows the user to select the priority of the alarm. Choices are 1, 
2 and 3. Priority 1 alarms are considered the most serious.
ALM PRINT Allows the user to enable or disable automatic printing during a 
physiological alarm condition. Choose ON to enable printing 
upon alarm. Choose OFF to disable printing upon alarm. See 
“Recorder (Optional)” on page 2-82 for details.
HR ALM HI Allows the user to set the upper limit of the HR alarm. (See “Heart 
Rate Alarm Limits” on page 2-27.)
HR ALM LO Allows the user to set the lower limit of the HR alarm. (See “Heart 
Rate Alarm Limits” on page 2-27.)