TransducerControlsandInspection
Operator's Manual 23
Disconnecting the Transducer
1.
Afterfreezingorturningo󰀨theultrasonicdiagnosticsystem,togglethelockingleverto
theunlockpositiontounlocktheconnectoroftheprobe.
2.
Pulltheprobeconnectorstraightoutasshowninthegurebelow.
1
2
3.3 Electrical Safety
TheelectricalleakagecurrenttestshouldbeperformedontheTEEtransduceraftertaking
itoutoftheboxandpriortoeachexam,alternatively,ifthebite-guardinspectiontestis
donepriortoeachexam,thentheelectricalleakagecurrenttestshouldbedoneyearlyata
minimum.
3.3.1 Electrical Leakage Current Test
Mindrayultrasoundsystemswithaccessoriesaredesignedtomeettherequirementsfor
patientsafetydescribedinIEC60601-1MedicalElectricalEquipment-Part1.General
RequirementsforSafety.Tomaintainpatientsafetyitisimportanttohavealowelectrical
leakagecurrentintheproduct.
Theendoscopeshafthasnoelectricallyconductingsurfaces,andiscoveredwithalayer
ofmaterial,whichpermitsneitheruidsnorelectricitytopassthroughit.Electricalsafetyis
maintainedforthetransducerbykeepingthismaterialintact.EachTEEtransduceristested
forelectricalisolationandleakagecurrentbeforeitisshippedtoacustomer.
WARNING
To avoid injury to the patient, do not use the transducer if the insulating material
has been punctured or otherwise compromised.
Measuring electrical leakage current should only be done by qualified personnel.
Take all necessary precautions to avoid contact with non-insulated parts that
have applied voltage.