CleaningandDisinfection
36 Operator's Manual
5.1 Before Processing
Thisstepistoremovetheultrasoundgelorothervisibledirt.
1.
Wearapairofglovestopreventinfectionthroughthewholecleaning,disinfection
process.
2.
Disconnecttheprobefromthesystem.Ifthesheathisused,takeothesheathand
disposeitasdirectedbythehospital.
3.
Wipeawaytheultrasoundgelorothervisibledirtonthesurfaceoftheprobebyusing
adamppieceofdisposablelint-freesoftclothortissue.
5.2 Cleaning
OnlythefollowingcleanersarerecommendedbyMindraytocleantheTEEtransducers.
Forthebiologicaleectivenessandthecorrectuseofthecleaners,seetheinformationof
thecleaner’smanufacturer
.
5.2.1 Compatible Detergents
Cleaner Active ingredient Concentration Applicable for
Cidezyme/Enzol Proteolyticenzymes < 5% P7-3T
P7-3Ts
P7-3TU
P7-3TE
P8-3Ts
P8-2Ts
P8-2TU
XP8-2TU
Neodisher
MediCleanforte
Trisodiumnitrilotriacetate 5-10%
2,2-Iminodiethanol
diethanolamine
1-2%
Prolystica2x
conc.Enzymatic
Presoak&cleaner
Ethanolamine 1-5%
Protease 0
.
1-1%
Ethoxylatedalcohol 1-5%
Polyalkyleneglycol 1-5%
Glycerine 1-5%
EMpower Proteolyticenzymes < 2% P7-3T
P7-3Ts
P7-3TU
P7-3TE
P8-3Ts
P8-2Ts
P8-2TU
Metrizyme Proteolyticenzymes < 2%
WIP’ANIOS
CLEAN’UP
Associationofsurfactants,
excipients
/ P8-2Ts
P8-2TU
TEEZyme EthoxylatedAlcohols 0.1-1% XP8-2TU
Proteolyticenzymes 0.1-1%
Amylase 0.1-0.5%
TEEZymeSponge Proteolyticenzymes 0.1-1%
5.2.2 Cleaning Procedure for TEE Probe
1.
Selectanappropriatecleanerincludecleaningwipesandsprays,milddetergents,
enzymaticcleanersandspeciallydesignedenzymaticsponges.