StorageandTransportation
Operator's Manual 41
6
Storage and Transportation
Whenallexaminationsforthedayhavebeencompleted,confirmthatthetransducer
isingoodconditionandstoreitinasuitableplacesothatthenextexaminationcanbe
conductedsmoothly.Makesurethetransducerisadequatelycleanedanddisinfectedprior
tostorage.
CAUTION
When transporting the transducer, do not allow any part of the transducer to
protrude beyond the case. Never store a moist TEE transducer in the shipping
case.
Use the tip cover to enclose and protect the distal end of the probe from being
damaged during transportation and storage.
6.1 Transducer
Fordetailedenvironmentalconditionsofstorageandtransportation,see"1.3 Ambient
conditions".
WhenthetransducerissenttoMINDRAYCustomerServiceDepartmentorsales
representativeforrepair,besuretodisinfectitandkeepitintheoriginalcarrying/
shippingcasetopreventinfection.
Sterilizethecarryingcaseasnecessary.
6.2 Tip cover
Thetipcoverenclosesandprotectsthedistalend/scanheadoftheendoscopefrombeing
exposedtomechanicalstrainduringtransportationandstorage.
CAUTION
To avoid damaging the transducer, the tip cover is a single use device. Discard after
use.