Instruction Manual MIT-SCAN-T3 Version 02/2021
Page 41
Error message
What to do?
Max. position
wrong
Repeat measurement at measuring site. Place probe
about 30 cm before the reflector.
Measurement
aborted
Movement of probe must be forward..
Electromagnetic
interfer.
Measurement signal was influenced by electromagnetic
interferences, e.g. by:
- High-voltage lines
- Underground cables
- Transformers, generators or other construction
machinery
Zero meas. invalid
Reference signal was influenced by extraneous metal.
If possible remove extraneous metal. Alternatively,
pass over reflector in other direction.
Incorrect
measurement
This error message appears if the course of the
measurement signal indicates that a rectangular foil
was not passed over along its long side. Repeat run
over rectangular foil along the long side.