CheckPoint 3 EN User Guide 49
P/N 340486-B
09/2017
New product
Creates a new, blank product. The following parameters can be set for a product:
TIP! If a new product will be very similar to an existing product, you might want to use the
“Copy Product” function.
When finished, a warning appears when exiting the function:
Tap Ye s to save the new product and return to the Products menu.
Product name A unique product name (up to 32 characters).
Product ID Unique product identification code (up to 32 characters).
Measure CO2 Select whether or not the measured CO
2
value should be
displayed on the measuring screen.
(Only applies to models with CO
2
sensor)
Display Balance Select whether or not the balance value should be displayed
on the measuring screen.
O2 high alarm limit
O2 low alarm limit
CO2 high alarm limit
CO2 low alarm limit
Setting of the O
2
and CO
2
alarm limits (0.1 - 100%).
Setting an alarm value to 0 will deactivate the alarm.