EasyManua.ls Logo

Mocon CheckPoint 3 - Calibration

Mocon CheckPoint 3
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DANSENSOR
66 User Guide EN CheckPoint 3
P/N 340486-B
09/2017
Calibration
In addition to the annual factory calibration, the device normally does not require further
calibration. If, though, the device measurements deviates more than +/- 3% when measuring
gas with a known CO
2
content, the calibration function should be used.
NOTE! Calibration requires for special skills and the access to certified calibration
gasses. Invalid calibration compromises measurement results.
Always calibrate 0% CO
2
before you calibrate 100% CO
2
.
For the 0% CO
2
calibration, 100% N
2
, technical air or atmospheric air can be used.
When using atmospheric air, please ensure that there is no increased or unstable
CO
2
concentration in the surrounding air, as this may result in erroneous measuring
at relatively high CO
2
measurements.
Carry out any such calibration far away from the CO
2
application area, and ventilate
the device thoroughly using atmospheric air by carrying out 5 measurements with
"clean" atmospheric air before calibrating.
When using gas from bottles, e.g. for the 100% CO
2
calibration, the calibration gas
must not be forced through the device.
Instead you must let the needle suck the gas from an overflow to which between 0.2
and 1.0 l/min is supplied (see below).

Table of Contents

Related product manuals