EasyManua.ls Logo

Mohr CT100B Series - Specifications; Electrical Specifications

Mohr CT100B Series
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A. Specifications
A.1. Electrical Specifications
Characteristic Specifications Notes
Reflected rise time, CT100B
10-90%
150 ps typ, 200 ps max 10 to 90%, into 50
Reflected rise time, CT100B
20-80%
100 ps typ, 150 ps max 20 to 80%, into 50
Reflected rise time, CT100HF
10-90%
100 ps typ, 130 ps max 10 to 90%, into 50
Reflected rise time, CT100HF
20-80%
80 ps typ, 90 ps max 20 to 80%, into 50
Jitter 15 ps max
(peak-to-peak)
< 1 ps rms typ.
Output impedance 50 ± 2%
Pulse amplitude 300 ± 10 mV Averaged value into 50
Pulse width Fixed: 1.000 2.000,
5.000 6.000,
20.00 21.00µs
Averaged values change
for short, medium, and
long cable length
settings, respectively
Pulse repetition time Fixed: 4.000 ± 0.02,
20.00 ± 0.02,
80.00 ± 0.02µs
Averaged values change
for short, medium, and
long cable length
settings, respectively
Sequential sampling rate Fixed: 250.0 ± 0.2,
50.00 ± 0.04,
12.50 ± 0.02 kHz
Averaged values change
for short, medium, and
long cable length
settings, respectively
Vertical scale 0.5 mρ/div
Vertical accuracy ±3% full scale < ±1% of measurement
typ.
Vertical position Any trace point is
movable to the center of
the screen.
Sampling efficiency, CT100B 50 to 90% into 50
Sampling efficiency, CT100HF 45 to 90% into 50
Noise ±5 mρ peak-to-peak
Input susceptibility ±1 A max into diode clamps
CT100B TDR Cable Analyzers Operator’s Manual 111