4. Operating Instructions
1. Press the blue MENU button. The Main menu appears.
2. Choose Settings Save Trace Default Name.
3. A dialog box appears. See Figure 4.31. Press the Show Keyboard menu option to display the
on-screen keyboard, or use a USB keyboard.
4. Enter the default name to use for saved traces.
5. There are also options in this dialog box to automatically append either a revision number
(e.g., save_001, save_002) or a timestamp to the default name.
6. Select OK to save the default name.
Figure 4.31. The save trace default dialog box.
4.9.4. Load a Trace (Cable Records)
A stored trace can be loaded back to the screen:
1. Press the FILE button. The File menu appears.
2. Select the Cable Scan Records option. A scrolling dialog box (see Section 3.10.1.1) appears
for the selection of a stored trace.
3. Use the M-FUNCTION knob to highlight the trace you wish to load.
4. Press the Select menu option to load the trace.
The stored trace now appears on the screen. It can be selected and manipulated the same as any
scanned trace. See Figure 4.32 in which the loaded trace (red, top) has been translated slightly
upward using the VERTICAL POSITION knob in order to improve visualization.
62 CT100B TDR Cable Analyzers Operator’s Manual