Mohr CT100B Series - Math Menu

Mohr CT100B Series
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3. Getting Started
3.9.2.1. Math Menu
The Math menu is a menu selection under both the SCAN button and the blue MENU button.
Math menu selections are shown in Table 3.2.
Table 3.2. Math Menu
Menu Selection Purpose/Action
FFT Tools
FFT Settings
Window Sets window type (Hann, Hamming, etc.).
size FFT buffer size.
Frequency Cut-Off Range FFT display frequency cut-off.
Hide Selected Trace Hides the selected trace.
Apply FFT Creates FFT trace from selected trace data.
View Live FFT Creates live FFT trace from live TDR trace.
Apply VSWR Creates VSWR (voltage standing wave ratio) trace
from a selected complex division or S
11
trace.
Set Base Sets base trace for complex division FFT trace.
Complex Division Creates complex division FFT: Divides selected FFT
by base FFT. Equivalent to uncalibrated S
11
trace
when base FFT is of a shorted test port.
Phase Creates phase angle trace from selected FFT trace.
Smith Chart Toggles Smith Chart display for selected complex
division FFT or S
11
trace.
SParam & Normalize Tools
S
11
Calibration Performs Open-Short-Load S
11
calibration and creates
S
11
trace.
Normalize Pulse Rise Time Sets the rise time to apply to the selected normalized
trace.
S
11
Options
Pre-Filter Options Changes the filter options that smooth the aberrations
of the base traces.
Post-Filter Options Changes the filter options that limit the noise of the
base traces.
Other S
11
Options
CT100B TDR Cable Analyzers Operator’s Manual 17