ELI 250/210__________________________________________________________________________
6-14
Channel Number: 6
WEP info (if used)
Once this setup is completed ECGs can be transmitted by pressing the “XMT” key from the main menu and
selecting the record you wish to send.
Transmission information can be viewed in the test software window.
When completed the UUT display will return to the ECG directory and a “T” flag will be set indicating proper
transmission. The transmitted ECG can be seen in the target directory (C:\UNIPRO) listed as a new file.
Safety Tests
Please refer to the following industry standard documents for information regarding safety testing.
ANSI/AAMI ES-1
IEC 60601-1
The following tests are performed per the above specifications:
AC Hi-Pot Patient to Mains
Earth Leakage
Chassis Leakage
Patient Source
Patient Sink