EasyManua.ls Logo

MSA ALTAIR 2X - Page 28

MSA ALTAIR 2X
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ALTAIR 2X
28
Use
GB
Bump Test for Devices with XCell Pulse Technology
Devices with XCell Pulse technology can be bump tested with a combination of
a pulse check and a flow check.
(1) Press the button for about 1 second.
PULSE? is displayed.
(2) Press the button again while PULSE? is displayed.
(3) Follow the instructions in section 3.4
Fig. 9 Bump Test for devices without XCell Pulse technology
0 ppm
Bump passed?
PASS ERR
Button Press?
PageMenu
Measurement
N
N
Y
Y
Wait for gas being applied
and gas reading reaching
threshold