3.9 Displaying Device Data
The Information page can be accessed by pushing the button for over
one second.
The following modes are displayed:
Bump test
Peak Reading
The Peak/Hi and Min/Low values can be
cleared.
When this page is displayed, press the button
to clear.
- CLR is displayed
- Value is deleted
Short Term Exposure Limit The STEL reading calculated since power-ON
displays. When this page is displayed, press
the button to clear.
- CLR is displayed
- Value is deleted
STL value is automatically reset to zero when
the device is turned ON.
Time Weighted Average The TWA reading calculated since power-ON
displays. When this page is displayed, press
the button to clear.
- CLR is displayed
- Value is deleted
TWA value is automatically reset to zero when
the device is turned ON.