The response time t90 will be increased up to 70 seconds by using the Sensor gard.
6.5 Sensor Response to Interferants
Interference factors may differ from sensor to sensor and with life time.
It is not advisable to calibrate with interference gases.
This table does not claim to be complete. The sensor might also be sensitive to other
gases.
Gas Sensor Interferant
Combustible wide variety of combustible gases and vapours
Ammonia [NH
3
] 100ppm
20ppm H
2
S
2ppm
Ammonia [NH
3
] 1000ppm
30ppm Cl
2
5ppm
200ppm H
2
S
120ppm
10ppm NO
2
6.5ppm
200ppm SO
2
-20ppm
Carbon Monoxide [CO]
100ppm NO
69ppm
100ppm H
2
40ppm
100ppm ethanol
4ppm
100ppm NO
2
-5ppm
Chlorine [Cl
2
]
1ppm Br
2
1ppm
2.4ppm ClO
2
0.55ppm
20ppm H
2
S
0.1ppm
10ppm NO
2
4.5ppm
0.25ppm O
3
0.11ppm
Hydrogen [H
2
]
300ppm CO
<60ppm
15ppm H
2
S
<3ppm
35ppm NO
10ppm
10ppm HCN
3ppm
100ppm C
2
H
4
80ppm
Hydrogen Chloride [HCl]
0.2ppm AsH
3
0.7ppm
5ppm Cl
2
< +/- 0.1 ppm
20ppm HCN
7ppm
20ppm H
2
S
60ppm
100ppm NO
45ppm
10ppm NO
2
< +/- 0.5 ppm