Table 14 Default Sensor Settings
Gas
(Code)
Thread
Type
4
TruCal
Range
Default
Display
Resolution
3
Unit
Default
Alarm
1
Default
Alarm
1
Default
AO
(mA)
Alarm
2
Default
Alarm
2
Default
AO
(mA)
Alarm
Action
Default
Span
Value
Default
Alarm
Min
Alarm
Max
Range
Min
Range
Max
Span
Value
Min
Span
Value
Max
Ammonia
(40)
Coarse
0-100
PPM
1 PPM 10 5.6 20 7.2
Increasing
Non-Latch
25 3 100 0-25 0-100 25 FS
1
Ammonia
(41)
Coare
0-1000
PPM
10 PPM 100 5.6 200 7.2
Increasing
Non-Latch
300 100 1000 190 1000 290 FS
1
Carbon
Monoxide
(10)
Fine 0-100 1 PPM 15 6.4 30 8.8
Increasing
Non-Latch
60 15 1000 0-10 0-1000 5
FS
1
Carbon
Monoxide
(11)
Fine 0-500 1 PPM 50 5.6 150 8.8
Increasing
Non-Latch
300 15 1000 0-10 0-1000 5
FS
1
Carbon
Monoxide
(12)
Fine 0-1000 1 PPM 100 5.6 300 8.8
Increasing
Non-Latch
400 15 1000 0-10 0-1000 5
FS
1
Carbon
Monoxide
H
2
Resistant
(14)
Fine 0-100 1 PPM 15 6.4 30 8.8
Increasing
Non-Latch
60 15 1000 0-10 0-1000 5
FS
1
Carbon
Monoxide
(36)
Fine 0-100 1 PPM 15 6.4 30 8.8
Increasing
Non-Latch
60 15 1000 0-10 0-1000 5
FS
1
Carbon
Monoxide
(37)
Fine 0-500 1 PPM 50 5.6 150 8.8
Increasing
Non-Latch
300 15 1000 0-10 0-1000 5
FS
1
4 Operation