eventofashort-termpoweroutage,thewarm-uptimeforcarbonmonoxideandethyleneoxidesensorswillbesignificantly
less.SeeTable10WarmUpTimestoestimaterequiredwarm-uptime.
Table 10 Warm Up Times
ReadingLevel <10ppm
1
<1ppm
Timeleftunpowered
Timetowarmup
1min ≤5min ≤5min
8hours ≤30min ≤30min
5days ≤30min 2-4hrs
1month ≤2hrs 6-10hrs
1
TheminimumalarmleveloftheCOsensorsis15PPMandtheETOsensoris1PPM.Belowthislevelwouldnot
causeanalarmonthedevice.
4.2 Settings
TheULTIMAX5000isatoolfreetransmitter.ThetwoEZtouchbuttonsonthefaceofthedisplaycanbeusedtonavigate
throughthemenustructure.Thebuttonsaredesignedforusewithfingerswitha“press”and“release”action,andworkbest
withoutgloves.
Table 11 Navigating through the menu structure
Button Function
leftbutton(↓) scrollingthrougheachmenu
rightbutton(→) selectingaparticularmenuoption
Changingavalue
1. Selecttherelevantoptionwith→.
The arrow disappears and the first digit to change appears underlined.
2. Scrollthroughthenumberswith↓.
3. Use→tomoveontothenextdigit.
When the arrow reappears, the value changing is finished.
Whenenteringanewvalue,rememberthefollowingtoensuresettingsaresaved:
UseSAVEbeforeexitingorsettingswillbelost.
UseBACKtogobacktopreviousscreen.
UseCANCELtogobacktomainsettingsmenu(i.e.BACK).
UseHOMEtoreturntogasreadingdisplay.
4.2.1 InstrumentSettings
Thefollowingsettingsaresavedtothedevicememoryandwillnotchangeifthesensortypeischanged.
1. ScrolltoSettings.
2. SelectInstrument.
US Ultima X5000 Gas Monitor 36
4 Operation

Other manuals for MSA ULTIMA X5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MSA ULTIMA X5000 and is the answer not in the manual?

MSA ULTIMA X5000 Specifications

General IconGeneral
Ingress ProtectionIP66/67
SensorsElectrochemical, infrared
CommunicationHART, Modbus
Target GasesCombustible gases, toxic gases, oxygen
ApprovalsATEX, IECEx, UL, CSA, SIL 2