1 SafetyRegulations
1.1 CorrectUse
TheULTIMAX5000GasMonitor,hereafteralsocalleddevice,isagasmonitorformeasuringtoxicandcombustiblegases
aswellasoxygen.Usingsensors,thedeviceteststheambientairandtriggersthealarmassoonasthegasexceedsa
specificconcentrationlevel.
WARNING!
Readthismanualcarefully.Thedevicewillperformasdesignedonlyifitisused,installed,andmaintainedinaccordance
withthemanufacturer'sinstructions.Otherwise,itcouldfailtoperformasdesignedandpersonswhorelyonthisdevicefor
theirsafetycouldsustainseriouspersonalinjuryorlossoflife.
WARNING!
Donotusesilicone-typelubricantsinassemblingthedeviceanddonotallowsiliconevaporstobedrawnintotheflow
systemwhileinoperation.Siliconecandesensitizethecombustiblegassensor,therebygivingerroneouslylowreadings.
UseonlygenuineMSAreplacementpartswhenperforminganymaintenanceproceduresprovidedinthismanual.Failureto
dosomayseriouslyimpairsensorandgasmonitoringperformance,alterflameproof/explosionproofcharacteristicsorvoid
agencyapprovals.
Failuretofollowthiswarningcanresultinseriouspersonalinjuryordeath.
ThisdevicecomplieswithPart15oftheFCCRules.Operationissubjecttothe-followingtwoconditions:
• devicemaynotcauseharmfulinterference,and
• devicemustacceptanyinterferencereceived,includinginterferencethatmaycauseundesiredoperation.
ThisequipmenthasbeentestedandfoundtocomplywiththelimitsforaClassAdigitaldevice,pursuanttoPart
15oftheFCCRules.Theselimitsaredesignedtoprovidereasonableprotectionagainstharmfulinterference
whentheequipmentisoperatedinacommercialenvironment.Thisequipmentgenerates,uses,andcanradiate
radiofrequencyenergyand,ifnotinstalledandusedinaccordancewiththeinstructionmanual,maycause
harmfulinterferencetoradiocommunications.Operationofthisequipmentinaresidentialareaislikelytocause
harmfulinterferenceinwhichcasetheuserwillberequiredtocorrecttheinterferenceattheuser’sexpense.
NOTICE
ThisisaClassAproductinaccordancewithCISPR22.Inadomesticenvironment,thisproductmaycauseradio
interference,inwhichcasetheusermayberequiredtotakeadequate-measures.
NOTICE
TheXCellsensorreferstothesensorportionoftheDigitalSensorthroughoutthismanual.
FCCWarningStatements
Changesormodificationsnotexpresslyapprovedbythemanufacturercouldvoidtheuser's-authoritytooperatethe
equipment.
IndustryCanada(IC)WarningStatements
TheinstallerofthisradioequipmentmustensurethattheantennaislocatedorpointedsuchthatitdoesnotemitRFfieldin
excessofHealthCanadalimitsforthegeneralpopulation;consult-SafetyCode6,obtainablefromHealthCanada'swebsite
5 Ultima X5000 Gas Monitor US
1 Safety Regulations