8 Appendix:Specifications
Table 19 Specifications
SensorOptions
CO H
2
S SO
2
CL
2
NO
2
H
2
S
500PPM
Hydrogen1000
PPM
NH
3
-100
PPM
NH
3
-1000
PPM
SO
2
-100
PPM
OperatingRange
2
-40°Cto
+60°C
-40°Cto
+60°C
-40°Cto
+60°C
-40°Cto
+60°C
20°Cto
+50°C*
-20°Cto
+40°C
-20°Cto+50°C* -30°Cto
+50°C
-30°Cto
+50°C
-20°Cto
+50°C
Storage -40°Cto
+60°C
-40°Cto
+60°C
-40°Cto
+60°C
-40°Cto
+60°C
-40°Cto
+50°C
-40°Cto
+50°C
-0°Cto+20°C -20°Cto
+40°C
-20°Cto
+40°C
-20°Cto
+40°C
ZeroDrift
1
<1%FSper
year
<1%FSper
year
≤1%FSper
month
<1%FSper
month
<1%FSper
month
<1%FSper
month
<1%FSpermonth <1%FSper
month
<1%FSper
month
<1%FSper
month
SpanDrift
1
<2%FSper
year
<2%FSper
year
≤2%FSper
month
≤2%FSper
month
≤2%FSper
month
≤2%FSper
month
≤2%FSpermonth ≤2%FSper
month
≤2%FSper
month
≤2%FSper
month
Noise
1
<2display
units
<2display
units
<2display
units
<2display
units
<2display
units
<2display
units
<2displayunits <2display
units
<2display
units
<2display
units
Repeatability
1
<±2% <±1% ±1% ±1% <±10% <±10% <±10% <1% <±15% <±15%
Resolution 1ppm 0.1ppm 0.1ppm 0.1ppm 0.1ppm 1ppm 10ppm 1ppm 10ppm 0.5ppm
T90 <9s <23s <6s <12s <60s <60s <185s <60s <300s <30s
T50 <30s <20s <40s <20s <20s <10s
Humidity 10-95%RH 10-95%RH 10-95% 10-95% 15-90% 15-90% 15-90% 15-90% 15-90% 15-90%
ExpectedSensorLife 5years 5years 5years 5years 2yearsin
theair
2yearsin
theair
2yearsintheair 5years 2yearsin
theair
2yearsin
theair
Warranty 3years 3years 3years 3years 1year 1year 1year 3years 1year 1year
EnablingTechnology XCell-non-
consuming
XCell-non-
consuming
XCell-non-
consuming
XCell-non-
consuming
Electro-
chemical
Electro-
chemical
Electro-chemical Electro-
chemical
Electro-
chemical
Electro-
chemical
TruCal
3
Yes Yes No No No No No No No No
8 Appendix: Specifications