Specification
DigitalSensor
H
2
S H
2
S500ppm CO
T20 5sec 4sec 4sec
T50 7sec 6sec 6sec
T90 18sec 30sec 15sec
T10(recovery) 67sec 35sec 54sec
Specification DigitalSensor(O
2
)
Tlowestalarmsetpoint 5sec
Thighestalarmsetpoint 5sec
Table 22 Operating and Storage Temperature Range
Storage
TemperatureRange
Operating
TemperatureRange
PressureLimits
min°C max°C min°C max°C
X5000Transmitter -40 +60 -40 +60 86kPa-108kPa
JunctionBox -50 +85 -40 +60 86kPa-108kPa
DigitalSensor(H
2
S,CO) -40 +60 -40 +60 86kPa-108kPa
DigitalSensor(H
2
S,500ppm) -40 +60 -40 +50 86kPa-108kPa
DigitalSensor(O
2
) -40 +60 -40 +60 86kPa-108kPa
Table 23 Accuracy Spec
Accuracy
DigitalSensor(H
2
S) ±3ppmor10%ofreading
DigitalSensor(H
2
S,500ppm) ±3ppmor10%ofreading(-20°Cto+40°C)
-20%ofreading(-20°Cto-40°C)
+25%ofreading(+40°Cto+50°C)
DigitalSensor(CO) ±6ppmor±10%ofreading(-10°Cto+60°C)
+15%ofreading(-10°Cto-20°C)
-30%ofreading(-20°Cto-40°C)
DigitalSensor(O
2
) ±0.5%O
2
(FM6340)
±0,2%(v/v)or±2,5%ofthemeasuringrange(EN50104)
93 Ultima X5000 Gas Monitor US
10 Appendix: General Certification Information

Other manuals for MSA ULTIMA X5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MSA ULTIMA X5000 and is the answer not in the manual?

MSA ULTIMA X5000 Specifications

General IconGeneral
Ingress ProtectionIP66/67
SensorsElectrochemical, infrared
CommunicationHART, Modbus
Target GasesCombustible gases, toxic gases, oxygen
ApprovalsATEX, IECEx, UL, CSA, SIL 2

Related product manuals