Hardware Instructions for Use
36
Power Up Test
Warnings
While running the Keypoint.NET application Power Up test, the following hardware
/equipment related warnings may appear on the screen:
Current Stimulator may be malfunctioning:
Problem occurred during the power up test of the stimulator. Exit the Keypoint.NET program,
and verify the cable connection.
Do not use the current stimulator when this message is shown; the stimulator might be
defective.
No Hardware detected:
Communication between PC and Hardware cannot be established. Exit the Keypoint.NET
program, and check:
- the USB cable connection between the computer unit and the Rear Panel of Main Unit;
- the power cable connection on the Rear Panel of Main Unit;
- the HS Link cable connection on the Side Panel of the Main Unit.