Hardware Instructions for Use
10
Safety Requirements
CAUTION This device is intended to be used by qualified medical personnel,
knowledgeable in the field of neurophysiology and neurophysiological assessment, as well as
in the use of the product / Keypoint equipment.
This device has been designed and tested in accordance with IEC 60601-1 Medical Electrical
Equipment, cf. the text on IEC 60601-1 further below in this section.
The device has been designed for indoor use at temperatures between +10°C and +35°C
(+50°F to +95°F).
The mains plug must only be inserted in a mains socket outlet provided with a protective
earth contact. It is forbidden to use extension cords. It is forbidden to use multiple portable
socket outlets (MPSO).
WARNING Any interruption of the protective earth conductor inside or outside of the
device or disconnection of the protective earth connector is likely to make the device
dangerous. Intentional interruption is prohibited. The protective earth / ground conductor
should be checked regularly.