XLTEK Protektor32 Instructions for Use
19
Phenomenon
Basic EMC
Standard
Immunity Test Levels – Professional
Healthcare Facility Environment
100 kHz repetition frequency
Conducted Disturbances Induced
by RF Fields
IEC 61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Table 9 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
Test
frequency
(MHz)
Band
(MHz)
Service Modulation Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385 380 –390 TETRA 400 Pulse
modulation
18 Hz
1,8 0,3 27
450 430 – 470 GMRS 460,
FRS 460
FM
± 5 kHz
deviation
2 0,3 28
704 – 787 LTE Band 13, 17 Pulse
modulation
217 Hz
0,2 0,3 9
810 800 – 960 GSM 800/900, TETRA
800, iDEN 820, CDMA
850,LTE Band 5
Pulse
modulation
18 Hz
2 0,3 28
870
930
1,720 1,700 –
1,990
GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
217 Hz
2 0,3 28
1,845
1,970
2,450 2,400 –
2,570
Bluetooth, WLAN,
802.11 b/g/n, RFID
Pulse
modulation
2 0,3 28