EasyManua.ls Logo

Nellcor N-200 - Page 86

Nellcor N-200
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
80
Excessive patient motion may be making it impossible
for the N-200 to find a pulse pattern. If possible, keep
the patient still; check whether the sensor is applied
securely and properly and replace if necessary; use
C-LOCK ECG synchronization; move the sensor to a
new site; use a sensor that tolerates more motion; or
set the N-200 for Mode 3.
The sensor may be damaged; replace it.
The patient’s perfusion may be too low to allow the
N-200 to measure saturation and pulse rate.
6. Patient module cannot be connected.
Use only a NELLCORC -13-200, C-20-200, or
C-13-200M patient module.
Connector pins may be bent; replace the patient
module.
7. SpO2 or pulse rate change rapidly; pulse amplitude
indicator is erratic. ECG waveform is noisy.
Check the patient.
Excessive patient motion may be making it impossible
for the N-200 to find a pulse pattern. If possible, keep
the patient still; check whether the sensor is applied
securely and properly, and replace it if necessary; use
C-LOCK ECG synchronization; move the sensor to a
new site; use a sensor that tolerates more motion; or
set the N-200 for Mode 3.
An electrosurgical unit (ESU) may be interfering with
performance:
Move the N-200, the patient module, and the cables
as far from the electrosurgical unit as possible.
Plug the N-200 and the electrosurgical unit into
different AC power supply.

Table of Contents

Other manuals for Nellcor N-200

Related product manuals