Troubleshooting
81
Move the electrosurgical unit ground pad as close
to the surgical site as possible.
The sensor may be damp or may have been reused too
often. Replace it.
If using a sensor extension cable, remove it and
connect the sensor directly to the patient module.
8. ECG LOST is displayed.
Check the patient.
If the ECG signal is provided through the patient
module:
There may be a loose, unplugged, or defective ECG
electrode or electrode cable.
Check the connections. If the problem persists,
replace the electrodes and/or ECG cables.
If ECG is provided through the patient module, the
module may be defective. Try another patient
module.
R-wave amplitude may be insufficient.
Check the ECG electrode position to verify proper
lead configuration.
Select (LEAD SELECT) limb lead with greatest
R-wave amplitude.
ECG signal provided via the rear-panel connector is
inadequate:
There may be a loose, unplugged, or defective ECG
cable.
Check the connections. If the problem persists,
replace the electrodes and/or ECG cables.