User's Manual
Oriel
®
LCS-100
TM
Small Area Sol1A

Table of Contents

Questions and Answers

  • V
    Virginia LeachAug 2, 2025
    What to do if Newport Oriel LCS-100 Laboratory Equipment lamp does not start?
    • K
      Kristin HallAug 2, 2025
      If the Newport Laboratory Equipment lamp doesn't start, several factors could be at play. First, ensure the power switch is turned ON and the orange light is lit brightly, also check the power cord. If it's not a power issue, verify the lamp is correctly installed following the installation procedure. Firm connections are important. It could also be a defective or old lamp. Try repeating the ignition sequence or replacing the lamp. Check that the lamp has a starter wire. If these steps don't work, it may be due to insulation breakdown or an interlock fault, in which case you should contact Oriel Instruments. Also, make sure that the top cover is in place.
  • A
    Alex LewisAug 7, 2025
    Why is there a loss of power in the output beam of my Newport Laboratory Equipment?
    • T
      troyholtAug 7, 2025
      A loss of power in the output beam from your Newport Laboratory Equipment can occur due to several reasons. The optics might be dirty, so cleaning them is a good first step. Alternatively, the optical assemblies could be damaged, requiring replacement. Also, if the lamp is old, replacing it might restore the power in the output beam.
  • J
    Jenny HallAug 12, 2025
    What to do if Newport Oriel LCS-100 lamp explodes?
    • R
      Robert CarlsonAug 12, 2025
      If the lamp in your Newport Laboratory Equipment explodes, it could be due to several reasons. The lamp might be defective, in which case you should replace it. It could also be an old lamp; replace lamps when stability becomes erratic, or output is less than 60% of original. Another cause might be a contaminated lamp envelope, so avoid handling the inside of the lamp reflector or lamp, and clean the lamp if needed. Also, ensure that the air vents are not blocked, needing > 1” clearance, < 40°C ambient air.
  • D
    Debra CookAug 17, 2025
    Why is there a loss of beam uniformity in Newport Laboratory Equipment?
    • A
      anthony09Aug 17, 2025
      A loss of beam uniformity in Newport Laboratory Equipment is typically caused by dirty optics. Cleaning the optics should resolve this issue.
  • A
    Alan SimmonsAug 22, 2025
    Why is the output unstable in Newport Oriel LCS-100 Laboratory Equipment?
    • D
      Daniel MartinAug 22, 2025
      If the output of your Newport Laboratory Equipment is unstable, it might be due to an old lamp. Replacing the lamp could solve the problem.
  • S
    Sara LarsenAug 27, 2025
    What to do if the shutter does not open in Newport Laboratory Equipment?
    • J
      Jeffrey DaughertyAug 27, 2025
      If the shutter on your Newport Laboratory Equipment does not open, it could be due to a damaged manual shutter or the shutter being mechanically jammed, in which case you should repair or replace the shutter assembly. It could also be a defective 71446 shutter drive circuit, so contact Oriel Instruments.

Summary

SAFETY

UV AND VISIBLE RADIATION

Hazards from high-intensity UV and VIS radiation exposure to eyes and skin.

LAMP EXPLOSION

Risks associated with lamp pressure and potential explosion due to internal strains.

OZONE

Hazards from ozone production by shortwave UV radiation and its effects.

ELECTRICAL SHOCK

Risks of electrical shock from high transitory voltage used for lamp ignition.

HEAT

Hazards associated with hot lamps and components during and after operation.

GENERAL WARNINGS

General safety precautions for operating and servicing the equipment.

GENERAL CAUTIONS

Precautions for proper use, maintenance, and handling of the equipment.

UNPACKING AND ASSEMBLING THE SYSTEM

FILTER INSTALLATION; REPLACEMENT

LAMP INSTALLATION; REPLACEMENT

OPERATION

IRRADIANCE ADJUSTMENT

Methods for adjusting irradiance using the mounting rod height.

TROUBLESHOOTING

Newport Oriel LCS-100 Specifications

General IconGeneral
BrandNewport
ModelOriel LCS-100
CategoryLaboratory Equipment
LanguageEnglish