retaining posts, and let it pivot back to its open position, as shown in Figure 11.
Figure 11: Lamp retaining clip and retaining posts
6. If an old lamp is in place, rotate it slightly and tilt it back so that the anode lead wire clears the
lamp seating plate, then lift it out of the lamp cage slowly, and set it down gently. For best control,
handle the lamp only by the ceramic mounting flange at the base of the reflector.
7. Remove the new lamp from its packaging. Grasp it by the ceramic base with the Anode wire
oriented upward for best maneuvering, as shown in Figure 12.
Figure 12: Lamp handling, insertion, and removal
8. Gently lower the lamp into the lamp cage. Rotate it slightly and tilt it back so that the anode lead
wire clears the lamp seating plate. Guide the reflector rim into the recessed lamp seat, using the
two white pegs as rests for guiding the lamp into place. Make sure the Anode wire is centered
within the cutout at the top of the lamp seating plate, as shown in Figure 13 and 14.
Retaining posts Retaining Clip