6 OPERATION
6.1 FUSE CHANGE FOR 220VAC
There are 2 fuses factory installed in the power cord jack at the rear panel, which are for 100
120VAC (nominal) operation. They are 1.5A type T (slo-blo).
For 200 – 240VAC (nominal) operation, you must replace these with the 0.8A type T (slo-blo) fuses
that are supplied in a bag shipped with the unit, as follows:
1. Unplug the LCS-100
TM
.
2. Use your finger or a small flat bladed screwdriver to slide out the fuse holder.
3. Flip down the outer wall a bit, which will raise the leading edge of the 2 fuses.
4. You may access the fuses with the aid of the screwdriver.
5. Replace them with the 0.8A type T (slo-blo) fuses that are supplied, flip the wall up and slide it into
place until it is flush with the connector.
Figure 17: Fuse replacement