8-4 WF1945B
8.2 Output voltage
z Output voltage (FUNCTION OUT)
Output range
Amplitude
Range
Resolution
Accuracy
DC offset
Range
Resolution
Accuracy
Amplitude and DC offset
limiting
Output impedance
Load impedance
Output connector
Others
10 V range / 1V range fixed, or automatic switchable
10 V range : 0 mVp-p to 20.000 Vp-p/open
1 V range : 0.0 mVp-p to 2.0000 Vp-p/open
10 V range : 1 mVp-p/open
1V range : 0.1 mVp-p/open
Continuous oscillation, external AM off,
, 1 kHz, rms value measured
10 V range : ± [0.7 % of amplitude setting (Vp-p) + 0.05 Vp-p]/open
1 V range : ± [0.7% of amplitude setting (Vp-p) + 0.01 Vp-p]/open
10 V range : ±10.000 V/open
1 V range : ±1.0000 V/open
10 V range : 1 mV/open
1 V range : 0.1 mV/open
DC mode, External AM off, external addition off
10 V range : ± (0.5% of DC offset setting [V] + 0.07 V)/open
1 V range : ± (0.5% of DC offset setting [V] + 0.01 V)/open
If output voltage exceeds the following value, the OVER light blinks
and the output may be clipped.
10 V range : 11 V/open
1 V range : 1.1 V/open
50 , unbalanced
45 and more, nominal
Front panel, BNC receptacle (FUNCTION OUT)
Output voltage can be set with high and low level buttons.
z SYNC OUT Output voltage (SYNC OUT)
Output voltage
Output waveform
Rise/fall time
Output impedance
Load impedance
Output connector
0/+5V (open)
2.5 ns
50 , unbalanced
45 and more
Front panel BNC receptacle (SYNC OUT)