EasyManua.ls Logo

NHP TemBreak PRO - INST Protection Only Setting

Default Icon
105 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning
41
TemBreak PRO B_BE-UM-001-EN V1.1.0
INST Protection Only Setting
The B_BE can be configured for INST protection only by selecting the INST only pre-defined time-current curve as follows: Refer to Protection Settings
Instantaneous Protection (INST) section for further detail on setting I
i
.
Action
Note / Illustration
1
Turn the MCCB to the OFF Position
Open the transparent flap in order to access Curve
adjustment dial.
2
Move the Curve adjustment dial using a PH1, PH2 or PZ2
size screwdriver.
Position the adjustment dial on the curve below that matches
the MCCB model:
B250 Curve 8
B400 / 800 / 1000 / 1250 / 1600 Curve 10
3
B800 / 1000 / 1250 / 1600 ONLY
Using a PH1, PH2 or PZ2 size screwdriver, rotate the I
R
adjustment dial to further adjust I
i
.
Refer to Protection Settings Instantaneous Protection (INST)
section for further detail on setting I
i
.

Table of Contents

Related product manuals