Chapter 4 Assembly
57
4
(7) Connect the control cable attached to the
DIH-M to the C-BOX.
1) Remove the cover on the left side of the
DIH-M.
2) Loosen the optical path switching section
fixing board mount screw.
3) Loosen the two optical path switching
section fixing screws and remove the clamp
from the inside of the optical path switching
section. (This frees the optical path
switching lever.)
4) Move the fixing board slightly forward with
the fixing screws attached, then tighten the
mount screw.
5) Replace the cover.
6) Loosen the zooming section fixing screw to
remove the clamp from the DIH-M.
7) Attach the blind cap provided.
(8) Connect the control cable attached to the
DIH-M to the C-BOX.
(9) Attach the hand switch for the
Epi-illumination shutter to the C-BOX.
Attach the hand switch fo
r
the Epi-illumination shutte
r
here.
Connect the control
cable attached to the
DIH-M here.
Fixin
g
screws for o
p
tical
path switching section
Mount screw for fixin
g
board for optical path
switching section
Cover
Fixing board for
optical path
switching section
Blind ca
p
Fixing screw for
zooming section