Chapter 6 Assembly
1 Assembly of Standard Set
29
6
2
Removal of Shipping Clamps
The stage top plate and focusing mechanism are
clamped for protection against vibration and shocks
during transportation. Remove these shipping
clamps with the hexagonal wrench provided.
• Stage Top Plate:
The top plate of the stage is retained in the
Y-axis direction by a plate fastened with two
bolts. Remove the bolts and the plate.
• Focusing Mechanism:
Remove the field lens unit to expose a plate
retaining the vertical movement of the focusing
mechanism. The plate is fastened with 3 bolts.
Remove the bolts and the plate.
3
Installing the Eyepiece Tube
Loosen the eyepiece tube clamp screw by hand and
place the eyepiece tube on the circular dovetail
mount. Tighten up the eyepiece tube clamp screw
by hand.
4
Connecting the Power Cord
Turn off the power switch of the microscope (press to the “
○
” position).
Connect one end (socket) of the supplied power cord to the AC inlet on the rear of the
microscope. Connect the other end (plug) to an AC electrical outlet with the ground conductor
(earth conductor).
Make sure that the power cord is securely connected.
Be sure to use provided (or specified) power cord. Use of other power cords may result in
malfunction or fire.
This completes the assembly of the microscope standard set.
Eyepiece tube
clamp screw
Shipping
clamps