Chapter 8 Troubleshooting
2 Operational
42
Binocular images not integrated.
Causes Corrective Measures
Interpupillary distance is not adjusted
correctly.
Adjust. (P.13)
Diopter is not adjusted correctly.
Adjust. (P.12)
Excessive eye fatigue.
Causes Corrective Measures
Interpupillary distance is not adjusted
correctly.
Adjust. (P.13)
Diopter is not adjusted correctly.
Adjust. (P.12)
Inadequate brightness or illumination.
Adjust brightness using the control dial.
(P.17)
Specimen does not move smoothly.
Causes Corrective Measures
Specimen holder is not securely fastened
to the stage.
Fasten securely. (P.31)