124
Specifications
7
7.1
Microscope (Ti-E or Ti-E/B) with TI-DH Dia Pillar Illuminator 100W
System configuration
Microscope, TI-DH Dia Pillar Illuminator 100W, and TI-PS100W Power Supply
Dimensions
260 (W) x 559 (D) x 729 (H) mm
Mass
26.5 kg
Optical system
Objectives: CFI60
Eyepieces: Field number 22
Nosepiece: Six sockets
Mechanical system
Focusing mechanism: Stroke: 10 mm
Focus knobs: Coarse focus: 300 μm per rotation (reference value)*
Fine focus: 50 μm per rotation
Extra fine focus: 6.25 μm per rotation
* When set to COARSE mode, the distance traveled by
rotating a focus knob will vary depending on how fast
the knob is rotated.
D-LH/LC
Precentered Lamphouse
LC
Input ratings: 12 VDC, 100 W
Lamp ratings: 12 V 100 W halogen lamp
Specified lamp model: Halogen lamp (OSRAM HLX 64623 or PHILIPS 7724I)
Average lamp life: 2000 hours
TI-PS100W
Power Supply
Input ratings: 100 to 240 VAC (±10%), 1.8 A, 50/60 Hz
Built-in fuse ratings: 250 V T4A
Output ratings: 12 VDC, 100 W
Maximum output current: 8.4 A
Electric shock protection class:
Class I
Remarks: UL listed product, GS approved
Operating conditions
Temperature: 0 to 40°C
Relative humidity: 85% RH max. with no condensation
Altitude: 2000 m max.
Pollution degree: Degree 2
Installation category: Category II
Indoor use only
Transport and storage
conditions
Temperature: -20 to +60°C
Relative humidity: 90% RH max. with no condensation