Chapter 3 Operation
3.7 Aperture Diaphragm Operation
65
Precautions on condenser and aperture diaphragm
Be sure to fully open the aperture diaphragm when performing phase contrast microscopy with the TI-C
Condenser Turret. If the aperture diaphragm is not fully open, the optical path will be blocked.
The aperture diaphragm of the ELWD-S condenser is used only for bright-field microscopy. This aperture
diaphragm is independent of the phase contrast annular diaphragm. Adjustment of the aperture diaphragm
will not affect the annular diaphragm.